Step
Loosen the M3 locking set screw on the end-cap (opposite end from display). See Figure 11.
1
Unscrew and remove the end cap.
Using thumb and forefinger squeeze the power connector at top and bottom to disengage the
2
locking mechanism then pull to disconnect. See Figure 11
Unplug the 24V wiring from the module
3
Loosen the two 24V power module retaining screws. The screws are captive.
4
Pull the 24V power module out of the transmitter.
5
The replacement 24V Power Modules are supplied with the "J2" Jumper in the NON-RTD
6
position. For 24V Transmitters utilizing RTD sensors, move the "J2" jumper to the "RTD"
position on the power module shown above.
Insert the replacement 24V power module into the transmitter. Reattach the screws and tighten
7
to 0,4 – 0,6 N-M (3.5 – 5.3 Lb-in).
Re-connect battery connector.
Honeywell recommends lubricating the end cap O-ring with a Silicone Grease such as Dow
Corning #55 or equivalent before replacing the end cap.
Screw the end cap back on and tighten the M3 locking screw.
8
Dispose of the Power Module according to local regulations
9
Revision 9
July 2011

Table 14 24V External power module

OneWireless XYR 6000 HLAI Transmitter R120 User's Manual
6.4. Replacing 24V external power module
Action
6. Maintenance/Repair
J2
37