Electrical Alignment
If the group prompt "LOCKOUT" has been configured for "CALIB", +VIEW", or "MAX" selection, you
will not be able to access PEN 1 group prompt for electrical alignment. Set LOCKOUT to "NONE".
Step
1
2
3
4
5
6
7
4
Table 3 Procedure for Setting the Chart Time Line
Press
Open the door, apply power, and wait for
the recorder to run its power up tests.
Until you see:
SET
UP
Upper Display
SET UP
Lower Display
LOCKOUT
Until you see:
FUNC
Upper Display
Lower Display
LOCKOUT
Until "NONE" is in the lower display.
or
ATTENTION
"LOCKOUT" selection in "NONE." Use
"NONE LOCKOUT" only when you intend
to make field adjustments to the
configuration
Until you see:
SET
Upper Display
UP
ALIGN
Lower Display
PEN 1
You will see:
FUNC
Upper Display
ALIGN
Lower Display
PEN 1 0
The pen moves to an internal stop then
backs away and stops.
to move the pen to the zero position on the
chart (inner position of chart).
or
ATTENTION You can position the pen
over 1200 steps. Those steps are relative
units only. Step 0.0 is the inner most
position on the chart, and step 1200 is the
highest circumference position.
Replacement Kit Instruction 43-DR-33-87
Action/Result
One of the following
NONE
CALIB
+ CONF
+ VIEW
MAX
Never leave the