General Maintenance: Maternal Blood Pressure Tests
FECG: Connect an FECG test body to the monitor's front panel
FECG test body can be the assembly leakage cable. Measure the voltage
breakdown.
pass
MECG: Connect an ECG test body to the monitor's front panel
MECG test body can be a shorted MECG cable or the assembly leakage cable.
Measure the voltage breakdown.
pass
IUP: Connect an IUP test body to the monitor's front panel
test body can be a SensorTip cable (wrapped in aluminum foil) or the assembly
leakage cable. Measure the voltage breakdown.
pass
MSpO
: Connect an MSpO
2
The MSpO
test body can be a MSpO
2
foil) or the assembly leakage cable. Measure the voltage breakdown.
pass
FSpO
: Connect an FSpO
2
The FSpO
test body should be an FSpO
2
aluminum foil. Measure the voltage breakdown.
pass
US: Connect an US test body to the monitor's front panel
test body should be an ultrasound transducer wrapped in aluminum foil.
Measure the voltage breakdown.
pass
US2: Connect an US test body to the monitor's front panel
test body should be an ultrasound transducer wrapped in aluminum foil.
Measure the voltage breakdown.
pass
120 Series Maternal/Fetal Monitor
2015590-001
fail
fail
fail
test body to the monitor's front panel
2
cable and sensor (wrapped in aluminum
2
fail
test body to the monitor's front panel
2
cable and sensor wrapped in
2
fail
fail
fail
ECG
input. The
ECG
input. The
UA
input. The IUP
MSpO
input.
2
FSpO
input.
2
US
input. The US
US2
input. The US
9-27