GE H
DRAFT
EALTHCARE
D
5460683, R
IRECTION
EVISION
Section 8-35
Replacement ECG Module
8-35-1
ECG Module
8-35-1-1
Manpower
1 person, 30 minutes
8-35-1-2
Tools
Phillips Driver
8-35-1-3
Removal Procedure
1.) Loosen the 1 screw and remove frame rear cover
2.) Loosen the 4 screw and dis-connect ECG cable.
3.) Remove LOGIQ S7 ECG MODULE ASSY
8-35-1-4
Installation Procedure
Parts to be installed in reverse order on removal.
8-35-1-5
Functional Check
Perform Diagnostics.
1.) Select "UTILITY" -> "Service".
2.) From "DIAGNOSTICS", select ECG Test.
3.) Verify test results shows PASS.
3
DRAFT (J
Figure 8-164 Remove frame rear cover
Figure 8-165 Remove frame rear cover
Chapter 8 - Replacement Procedures
25, 2014)
UNE
LOGIQ™ S7 E
/P
XPERT
RO
S
M
ERVICE
ANUAL
8-139