Section 8-35

Replacement ECG Module

8-35-1
ECG Module
8-35-1-1
Manpower
1 person, 30 minutes
8-35-1-2
Tools
Phillips Driver
8-35-1-3
Removal Procedure
1.) Loosen the 1 screw and remove frame rear cover
2.) Loosen the 4 screw and dis-connect ECG cable.
3.) Remove LOGIQ S7 ECG MODULE ASSY
8-35-1-4
Installation Procedure
Parts to be installed in reverse order on removal.
8-35-1-5
Functional Check
Perform Diagnostics.
1.) Select "UTILITY" -> "Service".
2.) From "DIAGNOSTICS", select ECG Test.
3.) Verify test results shows PASS.