GE Amersham Imager 680 Operating Instructions Manual page 241

Table of Contents

Advertisement

The software is non responding
If...
Transferring large amounts of files to a
connected computer during operation
can impair the system's performance.
Problems with tablet computer
connection
If...
The tablet computer screen turns white
when starting the software.
The software display on the tablet com-
puter screen has frozen after a long time
of inactivity.
No or few button presses are registered.
Amersham Imager 680 Operating Instructions 29271308 AB
8 Troubleshooting
8.3 Problems with the software
Then...
Avoid transferring during operation.
Then...
Make sure you are connected to the cor-
rect network on the tablet computer, that
is same network as Amersham Imager
680.
If the instrument has recently been
switched on it may take a few minutes
for the instrument to come online. The
wireless network can be detected before
the instrumented is fully connected.
Normally the instrument becomes online
within 5 minutes.
Restart the software on the tablet com-
puter.
Click the home button on the tablet
computer and restart the software.
Use gentle and distinct touches on the
tablet computer to control buttons in the
software. A light and short tap is best
registered by the software.