Fluorescence Trans-UV
Capturing with the Fluorescence image capturing method option Trans-UV is performed
using the UV Trans Tray (transparent tray) and a gel sheet.
Fluorescence Trans-UV sample preparation
Follow the instructions below to prepare the sample.
Step
1
2
3
4
Follow the instructions below to perform fluorescence capture using Fluorescence UV.
Step
1
2
Amersham Imager 680 Operating Instructions 29271308 AB
Action
Cut out a gel sheet slightly larger than the sample size.
Place the gel sheet on the UV Trans Tray (transparent tray).
Place the sample on the gel sheet.
Proceed with the capturing.
Note:
After use, wash the gel sheet with mild detergent, rinse with water then dry
well. A gel sheet can be reused about 20 times.
Action
In the Capture tab, select Fluorescence. The options available for Fluores-
cence capturing are displayed to the right on the screen.
Choose the UV.
5 Operation
5.5 Capture an image
5.5.4 Fluorescence capture
159