Chemiluminescence with
incremental exposure
Follow the instructions below to perform chemiluminescence capture with incremental
exposure time.
Step
1
2
3
4
Amersham Imager 680 Operating Instructions 29271308 AB
NOTICE
When using incremental exposure, images that are not selected
will be deleted at saving.
Action
In the Capture tab, select Chemiluminescence.
Result: The options available for chemiluminescence capturing are displayed
to the right on the screen.
Select Incremental exposure in the exposure drop down menu.
Set the number of repetitive exposures in the Expose field using the plus
and minus buttons.
Check the check box Stop when saturated if you wish the exposure to stop
when the pixel intensity of the accumulated exposed images reaches satu-
ration.
Note:
The Stop when saturated function overrides the set repetitive exposures.
5 Operation
5.5 Capture an image
5.5.2 Chemiluminescence capture
125