Step
2
3
4
5
Amersham Imager 680 Operating Instructions 29271308 AB
Action
If too many bands are detected, then use the Sensitivity minus button to
decrease the sensitivity.
If too few bands are detected, then use the Sensitivity plus button to increase
the sensitivity.
To make manual changes, select one detection channel from: Sample,
Markers, Blue, Green or Red above the image and select one or more rele-
vant options.
To create a new band, point to the position in the image.
Result: A new highlighted band appears.
To shift a band position, select the band and shift the band position up
or down with the Position arrow buttons.
To adjust a band width, select the band and adjust it with the Width plus
and minus buttons.
To delete a band, select the band and click Delete.
To delete all detected bands in the image, click Clear.
Repeat actions in step 3 for other detection channels, if relevant.
When you are done detecting bands click the Next button or select the MW
button in the workflow. The Previous button will bring you back to the previ-
ous step.
5 Operation
5.8 Analyze the image
5.8.3 Band detection
191