(2) Cable connections
1) Remove the skin-top spiral piece and lock nut of the X-ray interface cable.
2) As shown in the figure below, pass the X-ray interface cable through the X-ray interface box cabinet,
and loosely tighten the lock nut.
3) Pass the X-ray interface cable through the cable clamp, and connect the cable connector to the X-ray
interface PCB. Do not subject the cable to any stress and also, in order to prevent the excess length
from touching the devices on the circuit board, adjust the cable length and tighten the skin-top spiral
piece in such a way that no gap will be left as shown in the figure below.
Cable clamp
No gap to be
left here
4) Using a wrench, fasten the skin-top spiral piece and lock nut which were loosely tightened.
2. Installation procedures
Lock nut
Skin-top spiral piece
Skin-top spiral piece
- 27 -
Connectors