Replacing The Pca-Pb Board; Removing The Pca-Pb Board; Mounting The Pca-Pb Board - Canon CXDI-80C Service Manual

Table of Contents

Advertisement

3.3.4 Replacing the PCA-PB board

3.3.4.1 Removing the PCA-PB board

1) Remove the four screws (XB1-2300-606) securing the PCA-PB board.
2) Slide the PCA-PB board in the direction of the red arrow shown in the figure, draw the
detector cable connector part out through the hole in the base, and remove the board.
XB1-2300-606 screws (x 4)
PCA-PB board

3.3.4.2 Mounting the PCA-PB board

As a basic rule, mount the PCA-PB board by following the procedure for its removal in reverse.
Pass the detector connector part of the PCA-PB board through the hole in the base, and secure
the PCA-PB board using the four screws (XB1-2300-606).
Reference:
Tightening torque: 58.8 N-cm (6 kgf-cm)
4. Repair Guide
When the board is assembled, ensure that the detector cable
connector part protrudes from the base by about 4.5 mm.
- 10 -
Draw the detector cable connector part out
from the hole in the base.
AC/DC box base