System Installation Procedure - Canon CXDI-80C Service Manual

Table of Contents

Advertisement

6.2 System installation procedure

(1) Wireless connection
No.
Item
Installing the driver of
1
infrared data
communication unit
Connecting the infrared
data communication unit
2
to the image capture
computer
Connecting the wireless
LAN access point with
3
the image capture
computer
Setting the wireless LAN
4
access point
Installing the wireless
5
LAN access point
Turning on the power of
6
the wireless detector
7
Registration of detector
Connecting the image
8
capture computer with the
X-ray interface box
Connecting the X-ray
9
generator to the X-ray
interface box
Establishing the CXDI
10
Control Software NE
settings
Connecting to the
11
network and setting the
output destination
Aligning the detector and
12
X-ray tube exposure fields
13
Calibration
14
Self-diagnosis test
15
Imaging test
16
Final check
17
Cleaning
2. Installation procedures
Conditions and checkpoints
• The driver installation must be completed
successfully.
• Connect the unit to the USB port on the image capture
computer, and start the link Controller.
• This connection must be made through a hub.
• Turn on the power of the image capture computer, and
start the AP setting tool.
• Establish the settings for wireless communication.
• The access point must be installed outside the patient
environment.
• The distance between the detector and wireless LAN
access point must be less than 8 m.
• A fully charged battery must be installed properly in
the detector.
• The LED on the detector must light.
• The distance between the detector and infrared data
communication unit must be less than 30 cm.
• Care must be taken not to drop the detector.
• The detector must be registered in a location outside
the patient environment.
• It must be possible for wireless communication to be
performed normally.
• This connection must be made through a hub.
• Route the LAN cable in such a way that no excessive
weight will be applied to it.
• Route the X-ray interface cable in such a way that no
excessive weight will be applied to it.
• Set the items which are required.
• The center positions of the detector and X-ray tube
must be aligned.
• No errors must be displayed.
• No errors must be displayed.
• Transfer images to the printer and storage unit, and
check their image quality.
• Carry out the final checks as per the check sheet.
- 18 -
Reference items
Refer to "7.1.2
Connecting the
infrared data
communication unit".
Refer to "7.1.2
Connecting the
infrared data
communication unit".
Refer to "7.1.3
Connecting the
wireless LAN access
point".
Refer to "7.1.4
Establishing the
wireless LAN access
point settings".
Refer to "7.1.5
Installing the wireless
LAN access point".
Refer to "7.1.6
Registration of
detector".
Refer to "7.1.6
Registration of
detector".
Refer to "7.1.7
Connecting to the
X-ray interface box".
Refer to "7.3
Interfacing with the
X-ray generator".
Refer to "7.4 CXDI
Control Software NE
settings".
Refer to "7.4 CXDI
Control Software NE
settings".
Refer to "7.5 Image
Quality Checks".
Refer to "3.4 QC
Tool Button" in
CXDI Control
Software NE Setup
guide.
Refer to "3.4 QC
Tool Button" in
CXDI Control
Software NE Setup
guide.
Refer to "7.5 Image
Quality Checks".
Refer to "7.6
Post-installation
checks".