2.4
MULTlAXlS MEASUREMENT SYSTEMS
The system is designed to measure up to six independent axes with any combination of inter-
ferometers and retroreflectors. In general, the laser measurement concepts discussed above
apply equally to any multiaxis measurement system (see Figure 2-4).
BEAM
Figure 2-4. M ultiaxis Measurement System
The multiaxis system is similar to the basic measurement system shown in Figure 2-7 with the
exception that a beam splitter is introduced in the laser beam to provide a second axis of mea-
surement. The X and Y axes are completely independent and generate separate measure-
ment signals. These X and Y measurement signals are compared individually to the reference
signal by the electronics to provide displacement data for measurement or control applications.
Additional beam splitters can be introduced along with corresponding interferometers, retro-
refectors, and receivers to provide up to six measurement axes.
2.5
MEASUREMENT COMPONENTS
The
measurement
components
comprise
that
portion
of
the
Laser
Transducer System
used
to
generate,
direct,
and
detect
the laser
beam
and
consist
of
the
following
units:
5501A Laser Transducer
10780A Receiver
10702A Linear Interferometer (and Option 001 Windows)
10703A Retroref lector
10704A Retroref lector
10705A Single Beam Interferometer
10706A Plane Mirror I nterferometer
10700A 33% Beam Splitter
10701A 50% Beam Splitter
10707A Beam Bender
10710A Adjustable Mount
10711A Adjustable Mount