HP 5501A Operating And Service Manual page 240

Laser transducer system
Hide thumbs Also See for 5501A:
Table of Contents

Advertisement

Table
6-4.
English/Metric
Pulse
Output
Electronics
System
Troubleshooting
(Cont'd)
MULTI-AXIS
PROBLEM
1.
If
same
problem
exists
on
all
axes,
check
the
following
possible
error
sources:
a)
System
does
not
work
(control
LED'S
on
the10763A's
should
bedimly
lit
when system
iscycling).
SYMPTOM
CAUSE
ACTION
Upon sending
1. "Start" circuitry
1. Check wiring of
"start" pulse
malfunction
switch to send
system does not
"start" pulse.
start.
Must send a
2. Control not passed
2. Check switch and
start pulse!
between 10763A
jumper settings
cards (apparent by
on 1076314 card.
control LED lit o n
(Should corres-
only "master" card.)
pond to table
5-2, Section V.)
3. Large deadpath
3. Check OPC line
will result in long
on backplane
initialization
(see Table 4-1,
cycle (>20 seconds).
Section IV),
Each 1076314 must
using TTL Logic
initialize its own
Probe. This line
axis before
it
should be cycling
passes control to
high and low.
another 10763A.
4. If OPC is cycling
properly, prob-
lem is probably a
defective 10763A.
5. If OPC is not
cycling properly,
problem can be a
defective 10760A.
Overflow error
10760A counters trying
Switch normal/
is flagged after
to count i n wrong
reverse jumper on
moving the op-
direction (negatively)
10760A card.
tics
a
short
distance in the
positive direction.
Laser not pro-
1. System not
1. Disconnect con-
viding feedback
starting (see
troller and hook
information to the
previous page).
up freq. counter.
controller.
(See Table 6-4a, 6-4b)
2. Controller not
(Indicated by
recognizing output
2. Jumper 10763 for
the controller
pulses from elec-
testing (See
not appearing to
tronics.
table 6-4c).
receive data from
the laser elec-
3. Look at output
tronics.)
pulse count.
(See test a).