5.10.2
PROCEDURE FOR DEADPATH CONVERSION.
For the 10763A installation, deadpath
should be carefully measured and set in the 14 DEADPATH switches.
NOTE
Maximum Allowable Deadpath*
Resolution Selected
I
Normal
Extended
Metres
*Deadpath exceeding these amounts can not be compensated for
l nches
I
836
The deadpath in metres or inches must be adjusted and converted to binary numbers before
it can be set i n the 10763A binary DEADPATH switches. Use the following procedure to adjust
and convert the measured deadpath to the switch settings.
83.6
Calculate N from this formula:
N
=
( x ) ( K ) . ( R )
CLOCK RATIO
where:
N
=
number to be converted t o binary
X
=
deadpath
K
=
1.605 08 X 105 for X i n inches
=
10 pin
or
6.23023 pin
K
=
6.319 202 X
l o 6
for X i n metres
CLOCK RATIO
=
16 for 800
kHz clock selected
or
CLOCK RATIO
=
8 for 1.6
MHz clock selected
or
R
=
10 for times 10 resolution selected
or
R
=
1 for normal resolution