Table
6-3.
Binary
Controlled
Comparator-Based
System
Troubleshooting
(Cont'd)
c) Check indicator LEDs on the 10762A Comparator card.
d) intermittent measurement signal on one or more measurement axes may indicate
CAUSE
CHECK
ACTION
1 . Misalignment of
Reduce beam size by
If optics are
optical components
rotating to small
misaligned and
aperature on laser
beams do not
head. Place align-
overlap, refer
ment target on
to Section II
receiver lens. Move
for alignment
optics along entire
procedure.
measurement path and
verify both return
beams overlap on the
target along the
entire length of
movement.
2. Misadjustment of
Monitor 10780A test
Refer to 10780A
10780A Receiver
point on back of
manual for adjust-
Gain
receiver, voltage
ment procedure.
should lie in range
0.5 - 1.OV dc. If
voltage >I.lV,
receiver could be
saturated. If volt-
age <.02V, gain may
be too low.
3. Low power in
Use power meter, if
Rotate interfero-
frequency used
available, and meas-
meter 90 degrees
for measurement
ure relative power
and look for in-
of horizontal and
crease in test
vertical polarization.*
point voltage.
4. 10780A mal-
Look at output sig-
Troubleshoot
function, cable
nal from 10780 and
appropriate
intermittent, or
input of fast pulse
component
107648 malfunction
converter 107648.
'Use laser power meter filtered for 6328 Angstroms
(e.g., United Detector Technology MOX Meter (w/filter) ).