Remove the locking rings.
Remove the bearing rings.
Remove the shaft.
Lower the turn plate and remove it.
Check all springs for correct mounting and damage (Fig. 14).
Check that the dimension between the screw connection and the magnet housing is 49.7
mm. Adjust if necessary.
A new or inspected turn plate should be installed in the reverse order:
Reinstall the turn plate.
Reinstall the shafts.
Reinstall the bearing rings.
Reinstall the locking rings.
Connect the ground cable and K68.
Reinstall the supporting wheels.
Reconnect the KBATT plugs.
Reattach the side covers and the cover with the cassette compartment.
Set switch S10 in OFF position to charge the capacitor bank.
Reattach the system upper cover.
Connect the unit to mains power supply outlet and switch it on.
Reattach all covers.
Check the function of the turn plate by pressing one of the turn plate buttons and turning
the unit. Check also that the turn plate returns to the park position when the button is
pressed a second time.
Check the unit by releasing an exposure.