Siemens MOBILETT Plus HP Service Instructions Manual page 27

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

Display messages
Dis-
played
messages
CAL 28
CAL 29
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1
Siemens AG
Medical Solutions
Mode of
Cause
operation
Generator ON
An error was
during startup
detected during pro-
gram execution. The
program is inter-
rupted.
Generator ON
The single tank has
been connected
incorrectly.
SPR8-220.061.03
Rev. 03
Possible reasons and advice
Check that all supply voltages on
D1 are present (see page 110**).
If CAL 28 reoccurs or remains
displayed, the ROM containing the
program may be defective.
Replace the software (IC's I72, I66
and I74) on D1 (see page 110**).
If the error remains after replacing
the software, replace D1.
Enter the cause and the error
message "CAL 28" in the service
protocol.
The single tank is missing or the
cable harness has been incorrectly
connected.
Check that connection K9 on D1 is
connected.
Check for possible damage in the
main cable harness between D1
and the single tank. Measure K9:
13-4 and move the arm and single
tank while the connections are
being measured (see page 140**).
If damage is detected, replace the
main cable harness (both cables)
(see section "Replacement of the
arm cable harness").
Page 15 of 22
09.03
CS PS 24
2 - 15
MOBILETT Plus HP