Miscellaneous
Replacement of main switch element
•
Remove the side covers and the cover with the cassette compartment.
•
Remove the battery indicator (see section replacement of battery indicator).
•
Remove the four screws (two on each side) that fix the cable winder assembly
(pos. 1 / Fig. 2).
CAUTION
•
Pull out the cable winder assembly by approximately 10 cm.
•
Disconnect the switch element by turning the red lock counter clockwise (pos. 1 / Fig. 3).
Siemens AG
Medical Solutions
Sharp edge!
The brake handle could be damaged.
Lock the brake handle, otherwise it will be damaged by the sharp
upper edge of the cable winder assembly.
SPR8-220.061.03
Rev. 03
09.03
Page 7 of 8
MOBILETT Plus HP
CS PS 24
8 - 7
8