Siemens MOBILETT Plus HP Service Instructions Manual page 72

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

8 - 6
Fig. 9
Pull out the cable and rewind it to bring the tension back to the cable winder spring.
Repeat that six times.
Reinstall the upper plate with distance ring, plastic role and Y-plate.
Pull out the cable completely.
Test the cable winder and brake function.
NOTE
Reinstallation is performed in the reverse order.
NOTICE
Attach three new cable ties (pos. 1 / Fig. 4).
MOBILETT Plus HP
1
The cable winder should not rewind by itself.
The brake force is adjusted via the right screw.
nut clockwise -> stronger (pos. 1 / Fig. 9).
nut counterclockwise -> softer.
During reinstallation pay special attention to the shelf of the
power transmission and the rail of the cable winder assembly,
which must be under the shelf!
We recommend that you pull out the power transmission unit to
make reinstallation easier.
SPR8-220.061.03
Rev. 03
09.03
Miscellaneous
Page 6 of 8
CS PS 24
Medical Solutions
Siemens AG