Troubleshooting chart
This chart shows the sequence beginning with the generator start through to exposure
release and ends with the developed X-ray film. Potential errors are mentioned in the
function box.
Additional hints that may be helpful during troubleshooting are listed under A-E in the sec-
tion "Troubleshooting instructions".
MOBILETT Plus HP
Plug in mains cable
Main switch "C"
Battery indicator
right outmost LED
Yes
mains cable
Main switch "B"
Yes
Display illuminated
Yes
kV and mAs
buttons OK?
Yes
lamp can be turned
Yes
can be activated/
deactivated
Yes
indicator is ON
Yes
Siemens AG
Medical Solutions
active?
Plug out
No or partly
Collimator
ON/OFF
Turn plate
Green
Ready light
SPR8-220.061.03
Rev. 03
No
Charge batteries
Voltage supply
Program run,
see section "P04 - Display,
LED and buzzer test"
Connections
No
Button function
Defective lamp
No
Lamp button
remote control (option)
No
Turn plate buttons
Defective magnet
Charging time
Displayed message
No
not acknowledged
Page 1 of 2
09.03
CS PS 24
3
Cables
Fuse
Cables
Fuse
MOBILETT Plus HP
3 - 1
A
B
C
D
E