Siemens MOBILETT Plus HP Service Instructions Manual page 18

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

2 - 6
Dis-
played
messages
Use 10
CAL 12
CAL 13
* See section "Monitoring the kV and mA values"
** Wiring diagram X037E
Tab. 1
MOBILETT Plus HP
Mode of
Cause
operation
Generator ON.
The single tank tem-
The message
perature measured
does not inter-
is > 50°C or
rupt a running
< 8°C.
exposure.
During
Stored filament
preparation
value is lost.
Generator ON
The voltage supply
(60 V DC) to the fila-
ment circuit is
beyond its limit.
SPR8-220.061.03
Rev. 03
Display messages
Possible reasons and advice
The generator is overheated and
must cool down or the generator
was stored in a cold room and must
warm up prior to use.
Too high or too low ambient
temperature.
As soon as the single tank
temperature is < 47°C, check the
measurement circuit for the
temperature (see page 140**),
contact K9.
Check the voltage in the backup
battery (see page 110**). The
battery should have a voltage
> 2.8 V
Check that the battery jumper is
securely attached (see page
110**).
Make a new adaptation (see
section "P10 - Fast adaptation").
Check fuse U6 on D8 (see page
160**).
Check the filament supply voltage
+60 V DC (see page 160**).
Page 6 of 22
09.03
CS PS 24
Siemens AG
Medical Solutions