HT and single tank error diagnosis
Exposure values, 133 kV 5 mAs, HTPOS & HTNEG 2 V/div,
20 kV/div; 200 µs/div (only the end of the exposure is shown)
HTPOS on
D1
HTNEG on
D1
Fig. 6 HTsignal, arcing between plus and minus side (defective tube). Replace single tank.