HP CodeMaster XL+ M1722A User Manual page 93

Defibrillator/monitor
Table of Contents

Advertisement

Installation and Setup
Specifications
Paper Size: 50 mm by 30 m (100 ft).
Recorder Mode: Automatically documents events and ECG during defibrillation
episodes. The recorder can be configured to run in either real time or with a six-
second delay.
Frequency Response: 0.5 to 40 Hz or 0.05 to 150 Hz selectable.
Size and Weight
Dimensions: 7.9 in H by 11.8 in W by 15.6 in L (20 cm H by 30 cm W by 39.7 cm
L).
Weight: 24 lbs (10.9 kg). Includes external paddles, battery, and recorder paper.
Battery
Type: Rechargeable sealed lead-acid battery. 4 Ah, 12 V nominal.
Charge Time: 2 hours to 90% of full capacity. 18 hours to 100% capacity.
Repeated charging to less than 100% will reduce useful life of battery.
Capacity: 2 1/2 hours monitoring or fifty (50) full-energy discharges or 1 hour
monitoring and recording.
Battery Indicators: Illuminated LED indicates battery is charging. LOW BATTERY
message appears on monitor when limited battery capacity remains.
AC Input: 100 to 230 VAC +/-15%, 50 to 60 Hz.
External Pacemaker (Optional)
Current Pulse Amplitude: 30 mA to 200 mA.
Pulse Width: 20 msec.
Rate: 40 ppm to 180 ppm.
Modes: Demand or fixed rate.
A-13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Codemaster xl+ m1722b

Table of Contents