Environmental Storage Conditions: -20 to 70 deg C, 90% relative humidity for 24
hours at 65 deg C, 15,000 ft altitude.
Monitor
Inputs: ECG may be viewed through paddles or patient cable. Lead I, II, III, or
PADDLES selectable. Additional leads (avR, avF, AvL, V Leads) and PADS are
available. Monitor and recorder indicate
selected ECG source.
Lead Fault: LEADS OFF message and dashed baseline appear on monitor if a lead
becomes disconnected.
Common Mode Rejection: Greater than 100 dB measured as per AAMI standards
for cardiac monitors (EC13).
Display Size and Type: 5 inch (12.7 cm) diagonal CRT for 4 seconds of ECG data
on screen; non-fade, fixed trace. Scrolling trace is selectable.
Sweep Speed: 25 mm/sec nominal.
Frequency Response: 0.5 to 40 Hz.
Heart Rate Display: Digital readout on monitor from 15 to 300 BPM.
Heart Rate Alarms: Three configurable pairs of high and low heart rate alarm limits
from 20 to 280 BPM.
ECG Output: 1V/mV.
Patient Cable Length: 10 ft (3 m).
Thermal Array Recorder
Event Summary: Stores and prints 3 seconds pre- and 8 seconds post-critical event
data for up to 28 events. Data retained after unit is turned off.
Annotates: Time, date, HR, event marker, ECG mode, defibrillator mode, selected
energy, actual delivered energy, peak current, and patient impedance.
Speed: 25 mm/sec.