Monitoring

Monitoring
Monitoring
3
Figure 3-1
Lead Select button
ECG Size button
HR Alarm button
Review button
QRS Beeper Volume
ECG Output

The Monitor Control Panel

To monitor a patient's ECG with the CodeMaster XL+, perform the following steps.
1
2
3
3-6
Prepare the patient for ECG monitoring.
Turn the Energy Select control to the Monitor On position.
Press
Lead
to select the ECG source. The selected source appears in the
upper right corner of the display. For example, PADS appears on the display
when it has been selected.
If the message LEADS OFF or PADS OFF appears on the display, inspect the
m
electrodes, patient cable, leadwires, and associated connections. If the selected
ECG source is not connected, a dashed line will replace the normal ECG trace.