Monitoring

External Monitoring

Recorder Errors

The message CHECK RECORDER appears if an error occurs while recording. If this
message appears, check the recorder paper supply. This message may also appear if
the recorder door is open.
External Monitoring
3
The ECG output provides an analog 1V/mV ECG signal for connection to an
external monitor. Compatible external monitoring divider cables are listed in Table
3-8, "External Monitoring Cables," on page 3-11.
Table 3-8

External Monitoring Cables

1000:1 voltage
Divider Cable
Connector
Type
Six Pin
Eight Pin
Twelve Pin
The ECG output is also compatible with the interface to the HP central station ECG
input.
Do not use the ECG output to synchronize another defibrillator. (The ECG-In to
NOTE
ECG-Out delay is 35 milliseconds.)
The connection of external equipment may increase leakage currents. Always
CAUTION
request that local safety personnel verify that multiple connected equipment comply
with local regulatory standards before putting such equipment into service.
HP Part No.
M1782A
M14482A
M1783A
3-11