Troubleshooting
Troubleshooting
Troubleshooting the Pacer
Table 7-3

Pacer Messages

Message
PACER
FAILURE
PACER OUT-
PUT LOW
Pacer current
highlighted on
the display.
STOP PACER
NO PADS
PADS OFF
ATTACH
PADS
Cause
The unit detected a delivered current
error.
The pacer cannot deliver the
required current.
An attempt was made to change pac-
ing mode while pacing.
Pacer is on and one of the following
occurs:
No pads adaptor attached.
Paddles attached.
Internal paddles attached.
Pads adaptor cable is attached and a
pad is off.
Pads adaptor cable is attached.
Charg
is pressed and pads are
Pacer
not attached.
is pressed
and pads are not attached.
Possible Solutions
1 Change pads and make
sure pads adaptor cable is
properly connected.
2 Turn the Energy Select
switch to Off (Standby),
then turn back to Monitor
On.
3 Restart the pacer.
If
PACER FAILURE
hap-
pens again, use a back-up
pacer and call for service.
1
Stop the pacer.
2 Change the pads and
check the pads adaptor
cable for proper connec-
tion.
3 Restart the pacer.
Stop the pacer before you
change the pacing mode.
Attach pads for pacing.
Attach pad for pacing. If pads
are already applied to patient,
change to a new set of pads.
Attach the pacer pads.
7-5