Monitoring Electrodes - HP CodeMaster XL+ M1722A User Manual

Defibrillator/monitor
Table of Contents

Advertisement

Monitoring
Preparing the Patient
defibrillator connector slot. If the ECG leads cable falls off or is incorrectly
connected, the message LEADS OFF appears on the display.

Monitoring Electrodes

Using Pads
Standard pads (HP1749A) allow you to monitor through the pads for defibrillation
and synchronized cardioversion. If you wish to use the pads for pacing however, you
must attach separate electrodes for monitoring. To use standard pads, perform the
following steps.
1
2
3
4
Using Paddles
For an emergency evaluation you can monitor a patient's ECG through the paddles
electrodes when leads are not attached to the patient.
Do NOT use paddles to monitor the ECG during elective
WARNING
cardioversion procedures when the instrument is in synchronous
(SYNC) mode. Refer to Chapter 4, Performing Synchronized
Cardioversion, for detailed information on performing elective
cardioversion.
Attach pads as instructed on the pads package.
Attach the pads adaptor cable to the defibrillator.
Connect the pads to the pads adaptor cable.
PADS is the only selectable ECG source.
3-5