Table 15 2-Year Maintenance Worksheet - Sievers Innovox Online Toc Analyzer - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 8: Maintenance
Table 15: 2-Year Maintenance Worksheet — Sievers InnovOx Online TOC Analyzer
Item
GLS (clean and refill with DI water)
Oxidizer reagent
Dilution water
High pressure valve seals
High pressure valve seals
Pinch valve tubing
High pressure valve seals
Exterior sample pump tubing (optional
equipment)
Syringe lubrication
Reactor peristaltic pump tubing
Carrier Gas CO
Scrubber cartridge (optional
2
equipment)
High pressure valve seals
High pressure valve seals
Acid reagent
High pressure valve seals
Exterior sample pump tubing (optional
equipment)
Syringe lubrication
Reactor peristaltic pump tubing
Sample/reagent peristaltic pump tubing
Waste tubing
Carrier Gas CO
Scrubber cartridge (optional
2
equipment)
High pressure valve seals
High pressure valve seals
High pressure valve seals
Exterior sample pump tubing (optional
equipment)
Syringe lubrication
Carrier Gas CO
Scrubber cartridge
2
(optional equipment)
High pressure valve seals
High pressure valve seals
High pressure valve seals
Pinch valve tubing
Exterior sample pump tubing (optional
equipment)
NDIR Filter
Syringe lubrication
Reactor peristaltic pump tubing
Sample/reagent peristaltic pump tubing
Norprene Chemical waste tubing
Carrier Gas CO
Scrubber cartridge
2
(optional equipment)
GE Analytical Instruments ©2016
Operating Time
1 month
1 month, or as needed
1 month
1 month
2 months
3 months
3 months
3 months
3 months
3 months
3 months
4 months
5 months
6 months, or as needed
6 months
6 months
6 months
6 months
6 months
6 months
6 months
7 months
8 months
9 months
9 months
9 months
9 months
10 months
11 months
12 months
12 months
12 months
12 months
12 months
12 months
12 months
12 months
12 months
196 of 270