Step 13: Calibrate The Analyzer - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 4: Installation
lines when the Flush is complete. If not, press Flush All again. (If one or more of the reagent, dilution
water, or check standard lines are not flushed, see
11. When the flush is completed, press the Back button to return to the Menu screen.
12. Press the Back button.
13. When ready, press the Start Analysis button. Press Main to return to the Main screen. A graph of the
CO
peaks will be displayed. Each peak appears after its peak height is measured, and the measured
2
NPOC concentration appears at the top of the screen.
14. When the analysis is complete, the CO
the Mass Response axis is only 1.0 µg full scale, then refer to the Troubleshooting section of this manual.
Otherwise, proceed to the next step.

Step 13: Calibrate the Analyzer

Before placing the Analyzer into normal operation, calibrate the Analyzer in a manner suited to your application.
Consult
Chapter 5, "Calibration and Verification"
GE Analytical Instruments ©2016
Chapter 9,
"Troubleshooting.")
peaks will be displayed on the screen. If there are no peaks and
2
for details on calibration options.
104 of 270
DLM 68100-09 EN Rev. A