Changing The Alarms On Or Off Setting; Resetting Alarms; Printer Setup - GE Sievers Innovox Operation And Maintenance Manual

Toc analyzer
Hide thumbs Also See for Sievers Innovox:
Table of Contents

Advertisement

Chapter 6: Analyzer Operation

Changing the Alarms On or Off Setting

You can turn the alarms on or off, as needed. The state selected (On or Off) appears on the Alarm On/Off button.
1.
Select the I/O tab.
2.
Press the button for the Alarm Output that is to be reset (such as the Alarm Output 1 button), The
corresponding Alarm Output screen appears.
3.
Press the Alarm On/Off button, and then press either the On or Off button that appears.

Resetting Alarms

You can reset an alarm while it is being output by the Analyzer.
To reset an alarm
1.
Select the I/O tab.
2.
Press the button for the Alarm Output that is to be reset (such as the Alarm Output button), The Alarm
Output screen appears.
3.
Press the Reset button. The Analyzer resets the alarm. The alarm is still active and will be output the
next time the concentration exceeds the selected threshold value.

Printer Setup

You can print data from the Analyzer directly to a USB connected printer. The InnovOx is capable of
communicating to the following list of printers:
DeskJet 3050
DeskJet D1341
LaserJet Professional P1102w
OfficeJet 6000
We do not support the use of any other printer. To install the printer, it must first be disconnected from the
InnovOx when power cycling or rebooting the Analyzer. Once the InnovOx is on, connect a USB communication
cable (not supplied with Analyzer). Select the I/O menu and select "Printer Setup" option. Select the Printer
Disabled button to change it to Printer Enabled. Select the Search button to have the InnovOx search for a
connected approved HP Printer.
To print data
1.
Select the Data tab.
2.
Click the Print button.
GE Analytical Instruments ©2016
162 of 270
DLM 68100-09 EN Rev. A