Chapter 6: Analyzer Operation
3.
Press the Measure button to specify which value to measure: IC, NPOC, TOC, or TC.
4.
Press Advanced Setup
set to On or Off.
5.
If Check Standard is disabled, press the Check Standard button and then press the Enabled button that
appears to the right.
6.
Press the Range button. The range buttons appear to the right.
7.
Press a button for the specified range (such as Up To 1,000) or press Auto to have the Analyzer use the
Auto Range function. (For more information on the Auto Range function, see
8.
Press the Acid button to specify the percent of this reagent to use in the protocol.
9.
Press the Oxidizer button to specify the percent of this reagent to use in the protocol.
10. Press the Reps button to change the default setting for the number of measurements (repetitions) the
Analyzer will make for this stream.
11. Press the Rejects button to change the number of check standard measurements the Analyzer should
exclude from its calculations.
N
:
The number of rejects must be less than the number of repetitions selected for the check standard.
OTE
Rejects applies to the initial measurements of the check standard, (as shown in
Figure 41 Continuous Operationn page
the concentration of the check standard is different from the previously measured sample, so that
the average of the check standard measurements is accurate.
GE Analytical Instruments ©2016
Figure 49: Check Standard (Setup) Screen
Blank Correction to specify whether the blank correction feature should be
127. Use this feature to exclude initial measurements when
137 of 270
"Auto Range" on page
144.)