Technical Specifications Calomat 62F - Siemens 7MB2531 Operating Instructions Manual

Continuous gas analysis
Table of Contents

Advertisement

Gas path
Measuring cell body
Internal gas path (gas lines)
Leak-tightness
Gas inlet conditions
Permissible sample gas pressure
Sample gas flow
Sample gas temperature
Temperature of measuring cell
Sample gas humidity
Purging gas pressure
7.5

Technical specifications CALOMAT 62F

Design, enclosure
Weight (analyzer)
Electrical characteristics
Power consumption
For non-heated gas connection block
For heated gas connection block
Fuse values
For non-heated gas connection
– 100 ... 120 V
– 200 ... 240 V
For heated gas connection
– 100 ... 120 V
– 200 ... 240 V
Gas path
Gas connections
Gaskets (O-rings)
Sensor
Internal gas path (gas lines)
Leak-tightness
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
Compact Operating Instructions, 09/2018, A5E45779389002-01
Stainless steel 1.4571
Stainless steel 1.4571
Leakage loss < 1 μl/s
800 to 1100 hPa (11.6 to 16 psi) absolute
30 to 90 l/h (0.5 to 1.5 l/min)
0 to 50 °C (32 to 122 °F)
60 °C (140 °F)
< 90% RH (relative humidity), non-condensing
< 165 hPa
Approx. 25 kg (55 lbs.)
Approx. 30 VA
approx. 330 VA
T1L250V (F3/F4)
T630L250V (F3/F4)
T1L250V (F1); T4L250V (F2/F3/F4)
T630L250V (F1); T250L250V (F2/F3/F4)
Female thread 1/8"-27 NPT (according to ANSI B 1.20.1)
FPM (e.g. Viton) or FFPM (e.g. Kalrez)
Glass
Stainless steel 1.4571 or Hastelloy C22
Loss through leaks < 0.5 μl/s
Technical specifications
7.5 Technical specifications CALOMAT 62F