4.3

Electrical connections

Please observe the following for correct operation:
• Provide a disconnector unit in the building installation.
• Connect the enclosure of the analyzer to the equipotential bonding (see also associated
operating instructions or product manual).
• Use solid or stranded cables with a cross-section of not more than 2.5 mm² for wiring the
terminals.
• Provide all stranded conductors with end sleeves.
• Keep all connection cables within the analyzer short.
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
Compact Operating Instructions, 09/2018, A5E45779389002-01