Service and maintenance
6.2 Analyzers for operation in hazardous areas
6.2
Analyzers for operation in hazardous areas
6.2.1
Analyzer
Following maintenance or the retrofitting of parts associated with explosion protection, the
analyzer may only be started when an Ex specialist has ensured that the analyzer has the
essential features and meets the requirements for explosion protection and issues a certificate
to this effect and/or has attached a test symbol to the equipment.
When restarting, make sure that the device has sufficient protection from impact energy of over
2 Joules in the area of the windows.
Annual maintenance of the analyzer must be carried out to check the electrical safety and
reliable operation, in particular the tightness of the sample gas path within the analyzer.
Furthermore, a leak test must be carried out following each intervention in the gas path
(containment system). The leak test is described in section Leak testing of containment system
(Page 46)
The owner can extend maintenance intervals in individual cases if no negative effects can be
expected on wetted parts materials (gaskets in particular).
Further information on maintenance of the analyzer can be found in the relevant operating
instructions or product manual.
6.2.2
Ex px safety equipment
Impermissible repair of the device
• Repair must be carried out by Siemens authorized personnel only.
The diagnostics and elimination of faults is described in the Operating Instructions of the
Ex px safety equipment concerned.
6.2.3
Additional equipment
Maintenance and troubleshooting of the additional equipment is described in the appropriate
Operating Instructions.
54
WARNING
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
Compact Operating Instructions, 09/2018, A5E45779389002-01