Additional Equipment Manual release 07/2002 g g as anal as analy y sis g g g y y y ULTRAMAT/OXYMAT/CALOMAT 6 Gas analyzers of category II 2 G for use in potentially hazardous areas of zone 1...
Page 2
CALOMAT 6, OXYMAT 6, ULTRAMAT 6 Gas analyzers of category II 2 G for use in in hazardous areas of zone 1 Additional Equipment Manual A5E00058873-02 Release 07/2002...
Toute communication ou reproduction de ce document, toute ULTRAMAT, OXYMAT, CALOMAT, SIPAN exploitation ou communication de son contenu sont interdites, sont des marques déposées de Siemens. sauf autorisation expresse. D'autres dénominations utilisées dans ce document peuvent Tout manquement à cette règle est illicite et expose son auteur également être des marques déposées dont l'utilisation par...
English Contents Notes Description Application range External explosion protection 1.2.1 Pre-purging phase 1.2.2 Operating phase Internal explosion protection Further safety measures 1.4.1 Analyzer versions EEx p with "Compensation of losses resulting from leaks" 1.4.2 Analyzer versions EEx p with "Continuous purging" Technical specifications Installation site Explosion protection...
If you require further information, or if problems arise which are not dealt with in sufficient detail in this document, please request the desired information from your local or responsible Siemens branch. Additional Equipment Manual EEx p Order No.:...
English Description Application range The C 6F, O 6F and U 6F analyzers are used to measure individual components in gas mixtures. In conjunction with a certified EEx p safety device guaranteeing the pressurized enclosure, these analyzers can be used in hazardous areas of zone 1. The EEx p safety device guarantees defined pre-purging in the housing, and also the required pressurized enclosure during operation.
English Further safety measures The sample gas and reference gas lines must terminate outside the hazardous area at a point with no danger; sample gas lines can be returned to the sampling point. When connecting combustible or non-combustible sample gases into the hazardous area, both the sample gas inlet and outlet must be provided with a flame inhibitor.
English Technical specifications Installation site Danger zone Zone 1 and zone 2 Ambient temperature 5 ... 45 °C Humidity Dew point must not be fallen below during storage and operation Explosion protection Ignition protection type - Pressurized enclosure EEx p with compensation of losses resulting from leaks (analyzer versions 7MB251x-xxx0x-xAEx, 7MB201x–xxxxx–2or3xxx and 7MB211x–xxxxx–2or3xxx)
English Pneumatic data of the protective gas Housing volume: Approx. 50 l Input pressure (to supply the EEx p safety device): 0.2...0.4 MPa (2...4 bar) Protective gas: - Inert gas (e.g. nitrogen from a gas cylinder) - Air from an ex-free zone; prerequisite: free from oil and dust, particle size <40 µm, oil content <5 mg/m³...
English Differential pressure monitoring (only for analyzer versions EEx p "Compensation of losses resulting from leaks" The sample gas pressure must be supplied fail-safe to the analyzer (within the specified limits, see Section 2.3). If only an operationally safe supply is available, an additional differential pressure switch must be provided (see also point 5 in "Special conditions"...
English Assembly Connection of sample gas and protective gas (purging gas) Installation should take place at a vibration-free location where possible and protected from direct solar radiation. The analyzer and the EEx p safety device must be fixed according to their dimensional drawings and connected according to the assembly diagram in Section 3.2 or 3.3.
English Assembly diagram for the versions EEx p with "Compensation of losses resulting from leaks" Additional Equipment Manual EEx p Order No.: A5E 000 58873-02...
English Electric connections Warning Proper handling of the equipment by qualified personnel with observance of the 8 8 8 8 warnings is a prerequisite for safe operation. The regulations for installation of electrical systems in hazardous areas according to DIN VDE 0165 (EN 60079-14) and the standards therein must be observed. Failure to do so can lead to considerable damage, serious injury or even death.
English Commissioning Commissioning can only be performed properly when the person doing the commissioning is familiar with the contents of the manuals and operating instructions provided. The warnings in the individual manuals must be observed in particular. Connection to the EEx p safety device The analyzer must be connected to the EEx p safety device as shown in the assembly diagrams 3.2 and 3.3: Check the power supply...
English Additional equipment It is possible to add additional equipment/components to a mounting rail in the left hand part of the analyzer. The following points must be observed: - The mounting rail is approx. 250 mm long which limits the number of components which can be installed.
English 6.1.2 Recommended test setup for ULTRAMAT 6F compressed air, free of dust, oil and condensation Bl 1, Bl 2 = blanking plugs M = manometer (relative pressure) 2 = outlet of sample gas path DR = pressure regulator 4 = outlet of reference gas path V1 = shut-off valve (leak rate <...
English - Sample gas inlet 1 must be closed by a blanking plug. A shut-off valve (e.g. needle valve) of sufficient tightness must be connected between sample gas outlet 2 and the sample gas dissipation. - A pressure sensor (measuring range: 50 kPa, resolution :1 hPa) must be connected to the supply line.