Parts in gas path wetted by sample gas
Sample chamber
•
•
•
•
Further designs upon request
Gas inlet conditions
Permissible sample gas pressure
Purging gas pressure
Sample gas flow
Sample gas temperature
•
Unheated devices
•
Heated devices
Sample gas humidity
7.3
Technical specifications OXYMAT 6F
Enclosure
Weight
Electrical characteristics
Power consumption
•
Unheated devices
•
Heated devices
Fuse values
•
Unheated devices
– 100 to 120 V
– 200 to 240 V
•
Heated devices
– 100 to 120 V
– 200 to 240 V
Parts in gas path wetted by sample gas
Standard devices
Gas path made of
Gas bushing
tubing
Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
Compact Operating Instructions, 09/2018, A5E45779389002-01
Titanium or stainless steel 1.4571
Body
or Hastelloy C22
Lining
Tantalum or aluminum
Gaskets (O-rings)
Material according to specification
Window
CaF
600 to 1500 hPa (8.7 to 21.8 psi) absolute
< 165 hPa (2.4 psi) over ambient pressure
18 to 90 l/h (0.3 to 1.5 l/min)
0 to 50 °C (32 to 122 °F)
0 to 80 °C (32 to 176 °F)
< 90% RH (relative humidity), non-condensing or dependent on measuring
task
Approx. 28 kg (62 lbs.)
Field device
Stainless steel 1.4571
; without adhesive
2
Approx. 45 VA
Approx. 330 VA
1T/250 (F3; F4)
0,63T/250 (FG3; F4)
1T/250 (F1); 0.63/250 (F2; F3; F4)
0.63T/250 (F1); 2.5T/250 (F2; F3; F4)
Technical specifications
7.3 Technical specifications OXYMAT 6F
Titanium or stainless steel 1.4571
or Hastelloy C22
Tantalum or aluminum
Material according to specification
CaF
; without adhesive
2
Field device in hazardous area
--
59