APPENDIX A REPLACING A 7320 TRANSMITTER WITH AN AV400

Before making
any electrical
connections,
refer to the
Warnings on
page 37
Factory-fitted Links.
DO NOT REMOVE
Earth (Ground)
Stud on Case
(see Fig. 6.10)
Cleaner
Receiver
7320 100 Sensor
Notes.
1) The screens of the sensor's emitter, receiver and cleaner cables must be enclosed in yellow/green sleeving and
connected to the earth (groud) stud fitted to the transmitter case closest to the cable's entry point – see Fig A.1.
2) Dual input versions – connect each sensor to the correct input terminals (A or B) as indicated on the sensor's
cables.
3) Use the three-hole cable gland provided with the sensor for the sensor cables.
4) Fit a link between C3 and C5. This stops a false Cleaner Fail messages occuring, as the 7320 system cleaner
module does not generate the Cleaner Acknowledge signal used in the AV400 for the Cleaner Fail Diagnostic.
5) Remove the existing tag from the green wire and fit an eyelet tag. Connect the green wire (Cable 2) to the AV400
Transmitter earth (ground) stud.
60
Terminal Block C
Terminal Block A
Cable 1
Cable 2
(2 Wire)
(3 Wire) –
see note 5
Emitter
Terminal Block B
Terminal Block B
Sensor
Sensor
B
A
Flowcell Connections
B1
B9
Emitter 0V
B2
B10
Emitter +24V
B3
B11
Emitter Trigger –ve
B4
B12
Emitter Trigger +ve
B5
B13
Receiver +12V
B6
B14
Receiver 0V
B7
B15
Receiver Reference Signal
B8
B16
Receiver UV Signal
Terminal Block C
Sensor
Sensor
B
A
Flowcell Connections
C9
C3
Cleaner +12V
C10
C4
Cleaner Trigger
Acknowledge
C11
C5
C12
C6
Cleaner 0V
Cable
Colors
Blue (Cable 2)
Red (Cable 2)
Green (Cable 1)
Yellow (Cable 1)
Red
Blue
Green
Yellow
Cable
Colors
Red
Green
N/A
Blue