8 SENSOR MAINTENANCE

8.1 Scheduled Maintenance

Warning.
DO NOT
uses high voltages that can cause serious injury or
death.
Note. The emitter and receiver modules contain no
user-serviceable parts and are sealed in clean and dry
conditions at the factory. Opening them will lead to
degraded performance. See also the
The following maintenance schedule is a general guide only. The
systems are designed for a wide range of applications where the
nature of the sample can vary considerably, therefore it is
necessary to amend the schedule to suit the particular
installation and sample conditions.

8.2 Cleaning the Sensor

Routine maintenance is limited to cleaning out the flowcell
manually to remove accumulated fouling or sediment. The
flowcell must be cleaned prior to calibration (see Section 4) to
ensure that the Zero and Span standards are not contaminated
by organic matter that may be present inside.
50
open the emitter module. It
above.
Warning
8.2.1 Dismantling and Cleaning – Figs. 8.1 to 8.4
Warning. Isolate the transmitter and sensor
from the power supply before dismantling the sensor.
UV
Warning. The
contains a high intensity xenon strobe lamp that emits
ultraviolet (UV) radiation. This must NOT be viewed
with the naked eye and must NEVER be operated
while outside the sensor. Under normal operating
conditions, it is not possible to see the light source but,
if the sensor is dismantled with electrical power applied,
it may be possible to expose the eyes to the strobe
flash.
Notes.
• The emitter and receiver modules contain precision
optical
components
accordingly.
• The emitter module contains the power supply,
voltage control and lamp components. As a result, it
is heavy and requires extra support.
• Do not support the modules by the cable(s) entering
their enclosures.
• Ensure that the O-rings are removed with the screw
collars securing the cleaner, emitter and receiver
modules; it is possible for these seals to be left inside
the flowcell.
sensor
emitter
module
and
must
be
handled