Chapter 2. Instrument Setup
2.5.1a
Setting the Probe Zero (PROBE & CAL - ZERO) (cont.)
User — Working with a clean, uncoupled probe, this method automatically determines the probe zero offset from the probe face
before coupling, and only performs zeroing on user request. It then takes the backwall echo afterwards. It also works in critical
coupling conditions or with rough surfaces
3.
After making your choice, use the joystick (
64
Your Complete Source for
Testing Equipment. Since 1969!
) to move to another function, or press HOME to return to the Main Menu.