System Delivery Procedure
Inspect and Set Up the System
1. Verify all three positions of the brake pedal are functional:
Tab. 1
Brake Pedal Functions
Pedal Position
Up
Center
Down
Fig. 5:
2. Inspect the system for cosmetic damage. Verify there are no scratches or blemishes on
the system.
3. Verify filter is in place.
Fig. 6:
Pull out the filter located at the rear of the system and check that the filter screen is in
place.
4. Position the ultrasound system in the office or exam room and set the brakes.
Siemens
Condition
Front wheels locked straight.
All wheels pivot freely.
Front wheel brake engaged.
Brake Pedal
S2000 Air Filter -- Pull to remove
US15-101.812.01.02.02
05.10
Page 13 of 40
Ultrasound
13
0
ACUSON S2000