Verifying and Entering
Setup Data
2-8
Before the Series 1 can make measurements, it must have the proper
setup data entered into its memory. GE Panametrics has entered all or
most of the setup data for each probe before delivery; however, you
should verify that all the data is correct and entered into your unit.
Note: If you want to switch a probe from one channel to another,
you must re-enter some or all of the setup data as described in
this section.
Verifying setup data consists of three steps:
Verify that probes are properly activated in the Probe
Configuration Menu as described on page 2-9.
Verify that calibration data for all necessary probes is properly
entered in the System Calibration Menu as described on page 2-12.
Verify that applicable high and low reference values for the Series
1 measurement circuitry are properly entered into the Reference
Menu as described on page 2-21.
All high and low reference values are factory set and
normally do not need adjustment. However, the factory
may instruct you to adjust the high and low moisture
reference values. If necessary, you should do so as
described in Performing an MH Calibration/Adjustment
Test on page 1-20 of the Service Manual.
Use the following sections to properly verify and enter setup data.
IMPORTANT: While you are reviewing data, you should take the
time to record it on the Program Information List
supplied in Appendix A of the Startup Guide, or on a
separate sheet of paper.