3. Press the
key (below the Select prompt). The Start
Calibration? for prompt is displayed and values for the set
test gases are displayed (toggled) below the prompt:
Calibrate
Start Calibration?
O2 Zero Gas
1.0 %
Abort
If a blowback routine has been configured, a prompt is
displayed to start a blowback routine.
4. Press the
key (below the Continue prompt). A Zero
calibration starts and a status bar indicates calibration
progress:
Calibrate
O2
20.9 %
COe
0.06 %
Calibrating Zero
Abort
When the Zero calibration routine is completed, a Span
calibration is performed automatically and a status bar
indicates calibration progress:
Calibrate
O2
20.9 %
COe
0.06 %
Calibrating Span
Abort
CI/AZ40-EN Rev. A
Continue
When the Span calibration routine is completed, a prompt
Sensors Are Settling is displayed and a status bar indicates
progress:
Calibrate
O2
20.9 %
COe
0.06 %
Sensors Are Settling
Abort
5. The calibration pass status is displayed. If the calibration is
successful, press the
to Operator page. If the calibration fails, a prompt Calibration
Failed / Slope Too Low is displayed:
Calibrate
O2
20.9 %
COe
0.06 %
Calibration Failed
Slope Too Low
Exit
6. Press the
key (below the Exit prompt) and refer to
Operating instruction OI/AZ40-EN for fault-finding flowcharts.
key (below the Exit prompt) to return
29