ABB Endura AZ40 Manual page 19

Oxygen and carbon monoxide equivalent coe analyzer
Hide thumbs Also See for Endura AZ40:
Table of Contents

Advertisement

Item
Description
Instrument air supply to sensor assembly:
supply required: 350 to 700 ±10 kPa (50.0 to 100.0 ±1.5
psig)
the dew point at line pressure must be at least 10 °C (18 °F)
below the minimum local ambient temperature
A
at the plant site
maximum particle size in the air stream at the instrument
must not exceed 3 microns
maximum total oil or hydrocarbon content, exclusive of
non-condensables, must be as close as possible to 0 w/w %
or v/v %. – it must not exceed 1 ppm w/w or v/v under
normal operating conditions
B
Shut-off valve
C
2-Stage coalescing filtration (self-draining)*
D
Instrument air pressure regulator
3-Way valve (optional for maintenance purposes only,
E
not necessary for operation)
Aspirator suction pressure port:
F
pressure required at port:
–51.7 to –65.5 kPa (–7.5 to –9.5 psig)
Aspirator suction pressure gauge (Magnahelic)**:
G
pressure range: 0 to –69 kPa (0 to –10 psig)
H
Test gas port (sensor test gas inlet)
Probe filter / pressure gauge**:
I
— pressure range: 0 to 20 in H
Zero test gas (cylinder)***:
— mixed gas of O
2
J
— nominal 1 % O
2
CO range used
— must be certified for both O
Table 3 Key to pneumatic installation schematic
CI/AZ40-EN Rev. A
O (inch WC)
2
/CO/N
balance
2
/ CO to be 80 to 100 % of the
and CO content
2
Item
Description
2-Stage cylinder regulator for zero test gas
K
set to 1 bar (15 psig)
Span test gas (compressed air supply or cylinder)***:
— concentration of O
used
L
— compressed air supply may be used for a 0 to 25 %
O
range (recommended)
2
— cylinder gas must be certified for O
— compressed air line may be defined as 20.95 % O
2-Stage cylinder regulator for span test gas
M
set to 1 bar (15 psig)
N
Flowmeter, test gas line
Table 3 Key to pneumatic installation schematic (Continued)
*Use 2-stage filtration only – required efficiency for 0.01 micron
(particles and droplets, installed in order) 93 % and 99.99 %.
**Gauges are required only at setup. If gauges are fitted permanently, a
shut-off valve should be used to prevent leakage from the gauge.
***Avoid locations near sources of heat – ambient temperature must not
exceed 49 °C (120 °F).
— Zero test gas should be the test gas of lowest oxygen content.
— Span test gas should be the test gas of highest oxygen content.
— For maximum accuracy, combine the highest CO test gas (CO span)
with the lowest (1 % nominal) oxygen test gas.
— The oxygen span gas should have the zero CO content
(CO zero).
— The oxygen span gas may be air (20.95 % O
to be 80 to 100 % of the O
2
2
content
2
) – recommended.
2
range
2
19