Flow Meter Replacement - Philips EverFlo Q Service And Technical Reference Manual

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

4. Install the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
5. Install the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
6. Install the Filter Cover. Refer to the Filter Cover Replacement section for more information.

8.16 FLOW METER REPLACEMENT

To remove the Flow Meter
1. Remove the Filter Cover. Refer to the Filter Cover Replacement section for more information.
2. Remove the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
3. Remove the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
4. Remove the Tubing Connected to the Flow Meter.
5. Using pliers, loosen the two speed nuts from the threaded fitting on the back of the Flow Meter.
6. Remove the Flow Meter from the Front Cabinet.
To install the Flow Meter
1. Align the threaded fittings on the back of the supplied Flow Meter with the holds in the Front
Cabinet.
The Flow Meter must be mounted with the shaft for the Flow Meter knob at the top.
2. While holding the Flow Meter in place, secure the Flow Meter to the Front Cabinet by hand
tightening the two speed nuts.
3. Connect the pressure tubing to the fittings on the Flow Meter.
Page 80
NOTE
Figure 8-30: FLOW METER REMOVAL/INSTALLATION
LO TM
E
F
S
& T
VER
ERVICE
I
ECHNICAL
NFORMATION
1039055, VER. 06