8.6

O 2 QUICK COUPLER REPLACEMENT

To remove the O
Quick Coupler
2
1. Remove the Filter Cover. Refer to the Filter Cover Replacement section for more information.
2. Remove the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
3. Remove the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
4. Cut the tie wrap securing the tubing to the O
5. Remove the securing nut from the coupler and push the coupler towards the outside of the device
to remove.
To install the O 2 Coupler
1. Insert the O
the flat "D" portion of the Coupler is aligned with the flat "D" portion of the pass through hole.
2. Install the securing nut to the O
3. Install the tubing to the O
4. Secure the tubing to the O
5. Install the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
6. Install the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
7. Install the Filter Cover. Refer to the Filter Cover Replacement section for more information.
Page 49
Coupler through the side wall threads toward the inside of the device ensuring that
2
Coupler and tighten.
2
Coupler.
2
Coupler using a tie wrap.
2
LO TM
E
F
S
VER
ERVICE
Coupler.
2
& T
I
ECHNICAL
NFORMATION
1039055, VER. 06