8.14 COMPRESSOR REPLACEMENT

To identify the type of Compressor
1. Units equipped with the 0.38 Compressor can be identified by the "1098014" and "2450AE38"
numbers on the Thomas label. This model compressor is not rated for Trans Fill models nor any
units converted to TransFill capability.
2. Units equipped with the 0.44 compressor can be identified by "1020070" and "2450AE44" numbers
on the Thomas label. This compressor should be used in all domestic TransFill units.
Philips Compressor
To remove the Compressor
1. Remove the Filter Cover. Refer to the Filter Cover Replacement section for more information.
2. Remove the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
3. Remove the Rear Cabinet. Refer to the Rear Cabinet/Power Cord Replacement section for more
information.
4. Remove the Main PCA. Refer to the Main PCA Replacement section for more information.
5. Remove the blue hose running from the Compressor to the Sieve Canister Assembly.
6. Cut the two tie wraps holding the mounting rod in place.
7. Remove the mounting rod securing the Compressor Springs to the Front Cabinet.
Page 70
GSSE Compressor
Figure 8-23: MOUNTING ROD TIE WRAPS