20. Once the firmware is upgraded retest the unit for the customer complaint.
21. If the unit passes, then continue with the repair process and final test.
22. When all the features match, a single known configuration will show in the bottom box.
If not, the following message appears "There are no matching configurations" and the device
needs to be reevaluated as to its original equipment configuration.
23. Click the Upgrade button and then proceed with Steps 13 thru 17.
Feature
By Feature:
Select the model number, compressor type (.44 or .38) and solenoid type (Air or O2
pilot)
Clear Filters:
Allows you to clear the By Features section
Cancel Button:
Cancel Button stops the firmware upgrade operation.
MSP version:
MSP version of the firmware installed.
Page 108
LO TM
E
F
S
& T
VER
ERVICE
Definition
I
ECHNICAL
NFORMATION
1039055, VER. 06