6.3.7
COMPRESSOR TEST
This procedure verifies if the Compressor is working properly.
EQUIPMENT REQUIRED FOR OPTION 1
Item
Reference Part No.
1
McMaster-Carr: Qty 1
of Part #: 5121K851
2
McMaster-Carr: Qty 1
of Part #: 7781K33
3
McMaster-Carr: Qty
of 1 of Part #:
4089K13
4
McMaster-Carr: Qty
of 1 of Part #:
5439K19
5
McMaster-Carr: Qty 1
of Part #: 2974K278
(Same as 5463K653)
Page 25
Description
White Polypropylene Barbed Tube Fitting
Reducing Tee for 3/8" X 1/8" X 3/8" Tube
ID
Compact Plastic Needle Valve 3/8" Barb X
3/8" Barb Connections
Multipurpose Gauge +/-2% Mid-Scale
Accuracy 2" Dial, 1/4" NPT Male Bottom, 0
- 60 PSI
High Pressure Tubing, 1/8" ID, 3/8" OD,
1/8" Wall
Durable Nylon Single-Barbed Tube Fitting
Reducing Coupling for 1" X 1/2" Tube ID
White
LO TM
E
F
S
& T
VER