To install the Rear Cabinet/Power Cord
1. Thread the Power Cord through the hole in the Rear Cabinet. A set of bent needle nose pliers will
help with the installation of the power cord.
2. Connect the Power Cord connectors to the J3 and J4 locations of the Main PCA. Refer to Figure
8-15.
3. Slide the Main PCA into the grooves in the Front Cabinet.
4. Secure the Power Cord Wires against the Sieve Canister with two cable ties.
5. Place the Rear Cabinet onto the Front Cabinet.
6. Secure the Front and Rear Cabinets together by torqueing the four hex screws (35 in-lbs.) and
two phillips screws (25 in-lbs.).
7. Install the Inlet Filter. Refer to the Inlet Filter Replacement section for more information.
8. Install the Filter Cover. Refer to the Filter Cover Replacement section for more information.
Page 48
Figure 8-8: STRAIN RELIEF REMOVAL